Furnace Structure |
- Double-layer steel case with two cooling fans which ensure low surface temperature
- High purity fibrous alumina insulation is used as external refractory
- Included vacuum flange with 30 mm quartz window. The laser beam can penetrate the furnace through the window and sapphire thermal block.
|
Power |
- 15 KW max (75A air breaker required )
- 8 KW for normal working temperature
|
Voltage |
AC 220V three phase |
Max. Temperature |
|
Heating Elements |
MOSI2 heating element |
Normal Heating/Cooling Rate |
5°C /min |
Heating Zone Length |
- Three zones: total length: 610 mm (Z1 & Z3 = 195 Z2 = 220 mm)
- Constant temperature zone: 300 mm (if set up three zone at 1600°C)
|
Tube Size & Vacuum Flange |
- High purity Alumina Tube
- 60 OD x 54 ID x 1200mm L
- 60 mm stainless steel water cooling vacuum sealing flange with valve and meter is included
- special quick clamp flanges for easy installation
- 30mm Dia x 3 mm double side polished single-crystal window installed on the flange, which allows the laser beam to go through the tube
- In order to block heating radiation and make laser bean transparent, Two sapphire rods with two ends polished must be placed inside furnace ( please order separately )
- 16L/min water chiller is required for cooling flange
|
Temperature Control |
- Three digital controllers via SCR (Silicon Controlled Rectifier) power control with phase angle fired
- Each controller is 30 programmable segment for precise control of heating, cooling rate and dwells time
- Built-in PID Auto-Tune function with overheating & broken thermocouple protection
- Over-temperature protection and alarm allows for operation without attendant(s)
- +/- 1 ºC temperature accuracy
- Three B-type thermocouples (one for each zone)
- RS485 Communications Port
|
European Temperature Controllers (optional upgrade) |
- 1 heating channel
- 1 program of 8 or 24 segment
- Alarm Relay: Form 'A' and Form ‘C’ built-in
- Digital Communications: RS485
- Temperature Accuracy: ± 0.1ºC
- Recipes: Using a PC tool recipes can be created that can be used to change the operating parameters of the 3000 simply by selecting a new recipe via the HMI. This is very useful where multiple products are processed using the same controller but require different parameters to be set.
|
Power Cable & Plug |
- A 10 feet length 8-3 AWG heavy duty UL approved power cable is included
- Extra length cable can be ordered by click picture left
- The plug is not included, but you can order from your local hardware store according to female plug in your lab.
|
Dimensions |
1500 L x 550 W x 940 H, mm |
Net Weight |
300 kg |
Warranty |
One year limited warranty with lifetime support (Consumable parts such as processing tubes, o-rings, and heating elements are not covered by the warranty, please order the replacement at related products page.) |
Compliance |
|
Shipping Package Size After Crating |
47"x45"x45"(1200L x 1150W x 1150Hmm) |
Shipping Weight |
400 kg |
Warnings |
- The tube furnaces with alumina tube are designed for use under vacuum and low pressure < 0.02 Mpa
- Never use graphite crucible in the alumina tube.
- Attention: A two-stage pressure regulator must be installed on the gas cylinder to limit the pressure to below 3 PSI for safe operation.
- The gas flow rate should be limited to < 200 SCCM ( or 200ml/min )
- Vacuum limit definition for all alumina tube furnaces: * Vacuum pressures may only be safely used up to 1500°C
|